Ethicon Harmonic Generator 300
The Ethicon Harmonic Generator 300 System utilizes ultrasonic energy to enable hemostatic cutting and/or coagulation of soft tissue. The system consists of an ultrasonic generator, a foot switch, an optional hand switching adaptor, a hand piece, and a variety of open and minimally invasive instruments. Note: Throughout this manual “instrument(s)” refers to The HARMONIC Generator 300 System blades, ball coagulators, or coagulating shears.
All the HARMONIC instruments vibrate longitudinally at 55.5 kilohertz. This ultrasonic vibration at the blade enhances its cutting ability. The same vibration seals small vessels with coagulated blood and tissue proteins. Hemostasis occurs when tissue couples with the instrument. This coupling causes collagen molecules within the tissue to vibrate and become denatured, forming a coagulum.
Ethicon Harmonic Generator 300 Applications
Ethicon Harmonic Generator 300 Features
- Hand/Foot Activated
- Improved Power Delivery
- Built-In System Diagnostics
- Easy-to-Read Control Panel
- Dual Foot Switch Receptacle
- State-of-the-Art Digital Core “Smart” Chip in handpiece defines Output Power Curve
It includes: Ethicon Harmonic Generator 300 (GEN 04), Aquiline Footswitch (FSW01), Ethicon Harmonic Scalpel (V92665)
Model Number / Product Code: GEN300 / GEN04
Voltage: 100 – 240 VAC
Product features
Product features
Made in highest quality, surgical grade stainless steel with satin finish to reduce glare. The instrument can be sterilized and reused.
Materials and care
Materials and care
PRODUCTS
This reprocessing instruction applies to reusable surgical and endoscopic instruments supplied by MEDLENE Medical.
GENERAL
Reusable surgical instruments from MEDLENE Medical can be used by medical professionals for a surgical procedure and can be reused after qualified reprocessing.
The professional user selects the appropriate instruments according to the intended function, the tissue to be manipulated and the anatomical structures.
WARNINGS AND PRECAUTIONS
MEDLENE Medical instruments must be cleaned, disinfected,
and sterilized before first use. Protective caps and transport packaging must be
removed beforehand.
A complete functional check must be carried out before each use. Improper use and overstraining due to twisting / levering can lead to breaks and permanent deformation.
Do not use metal brushes or abrasives, as there is a risk of corrosion due to surface damage. The safe
combination of instruments with each other or of instruments with implants must be checked by the user before clinical use.
Be careful when handling sharp tips and cutting edges - risk of injury.
In case of patients with Creutzfeldt-Jakob disease (CJK), suspected CJK or possible variants of this disease, the applicable national regulations regarding the preparation of instruments must be applied. Never leave instruments in disinfectant solution for too long. Follow the instructions of the respective manufacturer.